Instructions
4.Advanced User Guidebook 
56 
SPI Decoding menu list: 
MENU 
SETTING 
INSTRUCTION 
Bus Type  SPI  Set bus type of decoding as SPI. 
Configure 
SCLK 
Select the clock edge to match the signal,
sample the 
SDA data on the rising or falling edge of the clock. 
Time Out 
Set the minimum time that the clock (SCL) signal must 
be idle before the oscilloscope starts to search for the data 
(SDA) on which to trigger. The range is 30 ns to 10 s. 
Data Bits 
Set the data width of each frame to match the signal. It 
can be set to any integer between 4 and 32. 
Bit Order 
Select LS First or MS First to match the signal. 
Display 
Format 
Binary 
Decimal 
Hex 
ASCII 
Set the display format of the bus. 
EventTable 
ON 
OFF 
Select "ON" to display the event table. 
Save 
EventTable 
If a USB storage device is currently connected to the 
instrument,  save  the  event  table data 
in a .csv 
(spreadsheet) formatted file on the external USB storage 
device. 
4.  CAN Decoding 
To decode CAN signal: 
(1) Connect the CAN signal to the Signal Input Channel of the oscilloscope. 
(2) Adjust to the proper time base and voltage division. 
(3) In trigger menu, select Bus trigger, and select bus type as CAN, set parameters 
based on the characteristics of the signal, trigger the signal correctly and obtain 
stable display. Refer to "CAN Trigger" on page 51. 
(4) Push the Decode  button  on the front panel.  Select  bus type as CAN.  set 
parameters based on the characteristics of the signal. When the parameters are set 
correctly, the information carried by the signal will be displayed.  
Tip: If there are repetitive menu items in both trigger menu and decoding menu, you 
can set anyone of them, the other will be changed synchronously.  
Decoded information interpretation: 
Information 
Abbreviation 
Background 
Identifier 
I, ID, or do not display 
Green 
Overload Frame 
OF 
Green 