Instructions
4.Advanced User Guidebook 
51 
Time Out  Time out 
Set the minimum time that SCL must be idle, that is a 
period of SCL, available range 100ns-10s. Time out 
means SCL keeps idle for a specified time before 
oscilloscope starts to search for the data(SDA) on 
which to trigger. adjust M knob or click   to set 
time out, press     panel button or click 
 move cursor to choose which digit to be set. 
ClockEdg
e&Data 
Clock Edge 
Set Edge Clock as Rising edge or Falling edge. Means 
sample the SDA data on the rising edge or falling edge 
of the clock. 
Set the number of bits of the serial data character 
string. It can be set to any integer between 4-32. adjust 
M knob or click   to set Data Bits. 
Set the number of the data bits, ranges from 0-31, 
adjust M knob or click   to set Current Bit. 
Set the value of the current data bit as H,L or X (H or 
L). 
Set all the data bits to be the specified value in Data. 
Data Bits 
Current Bit 
Data 
All Bits 
Mode 
Holdoff 
Auto 
Normal 
Single 
Acquire waveform even no trigger occurred 
Acquire waveform when trigger occurred 
When trigger occurs, acquire one waveform then stop 
4.  CAN Trigger  
CAN (Controller Area Network) is a serial communication protocol of the ISO 
international standardization. 
By using the CAN bus trigger, you can trigger on Start of Frame, Type of Frame, 
Identifier, Data, ID & Data, End of Frame, Missing Ack, or Bit Stuffing Error. 
You need to specify the signal source, trigger signal type, sample point, and signal 
rate of the CAN signal. 
In  CAN  bus trigger mode, the trigger setting information is displayed on bottom 
right of the screen, for example,  ,indicates that trigger 
type is CAN, CH1 trigger level is -126 mV. 
CAN Trigger menu list: 