4.Advanced User Guidebook 
50 
Ack Lost 
Trigger when SDA data is high during any 
acknowledgement of SCL clock position. 
Address 
Trigger on the read or write bit when the preset 
address is met.   
Adr
For
mat 
Addr 
Bits 
Set Address Bits to be “7”
、
“8”or“10”. 
Set address according to the preset address bits, 
address range is 0-127, 0-255, 0-1023 respectively. 
Set Data Direction to be Read or Write. 
Note: The set is not available when Address bits is set 
to “8”. 
Addr
ess 
Dire
ction 
Data 
Search for the preset data value on SDA and trigger on 
the dump edge of SCL of the last bit of the data area. 
DatF
orma
t 
Byte 
lengt
h 
Set data byte length, available range 1-5 bytes. Adjust 
M knob or click   to set byte length. 
Select the data bit, ranges from 0 to (byte length*8 -1). 
Set data to be H, L or X (H or L) 
Set all the data bits to be the specified value in Data 
Curr
entB
it 
Data 
All 
Bits 
Addr / Data 
Trigger when Address and Data conditions are met at 
the same time . 
Mode 
Holdoff 
Auto 
Normal 
Single 
Acquire waveform even no trigger occurred 
Acquire waveform when trigger occurred 
When trigger occurs, acquire one waveform then stop 
3.  SPI Trigger 
Trigger on the specified data when the timeout condition is meet. When using SPI 
trigger, you need to specify the SCL and SDA data sources. 
In SPI bus trigger mode, the trigger setting information is displayed on bottom right 
of the screen, for example, ,indicates that 
trigger type is SPI, CH1 trigger level is 0.00mV, CH2 trigger level is 0.00mV. 
SPI Trigger menu list: 
MENU 
SETTING 
INSTRUCTION 
Bus Type 
SPI  Set vertical channel bus type as SPI trigger. 
Source 
CH1 
CH2 
Set CH1 as SCL or SDA. 
Set CH2 as SCL or SDA. 