Instructions
4.Advanced User Guidebook 
43 
windows, trigger source is CH1, polarity is positive, 0.00mV the differential between 
up level and low level threshold. 
Windows Trigger menu list: 
MENU 
SETTING 
INSTRUCTION 
Single 
Mode 
Windows 
Set vertical channel trigger type as Windows trigger. 
Source 
CH1 
CH2 
Select CH1 as the trigger source. 
Select CH2 as the trigger source. 
Threshold 
Up Level 
Low Level 
Adjust the M knob or click   to set the up level 
threshold. 
Adjust the M knob or click   to set the low level 
threshold. 
Condition 
Polarity 
Positive Polarity, the oscilloscope triggers on the 
positive Windows pulse.   
Negative Polarity, the oscilloscope triggers on the 
negative Windows pulse.   
Enter: Triggers when the trigger signal enters the 
specified trigger level range. 
Exit: Triggers when the trigger signal exits the 
specified trigger level range. 
Time: Specify the hold time of the input signal after 
entering the specified trigger level. The oscilloscope 
triggers when the accumulated hold time is greater than 
the windows time. Available range is 30ns-10s, default 
100ns. 
Mode 
Holdoff 
Auto 
Normal 
Single 
Holdoff 
Reset 
Acquire waveform even no trigger occurred 
Acquire waveform when trigger occurred 
When trigger occurs, acquire one waveform then stop 
100 ns - 10 s, adjust M knob or click   to set time 
interval before another trigger occur, press     
panel button or click   to move cursor to choose 
which digit to be set. 
Set Holdoff time as 100 ns 