Instructions
4.Advanced User Guidebook 
36 
In normal mode,  the  Horizontal Position and  Horizontal Scale  knobs  are used to 
adjust the horizontal position and time base of the Main window. 
In wave zoom mode, the Horizontal Position and Horizontal Scale knobs are used 
to adjust the horizontal position and time base of the Zoom window. 
Note:  
About the gestures for touchscreen, see "Touchscreen Controls " on P16.   
How to Set the Trigger/Decoding System 
Trigger determines when DSO starts to acquire data and display waveform. Once 
trigger  is  set correctly,  it  can  convert the unstable display to meaningful 
waveform. 
When DSO starts to acquire data, it will collect enough data to draw waveform 
on left of trigger point. DSO continues to acquire data while waiting for trigger 
condition  to occur. Once it detects  a  trigger it will  acquire enough data 
continuously to draw the waveform on right of trigger point. 
Trigger control area consists of 1 knob and 2 menu buttons. 
Trigger Level: The knob that set the trigger level; push the knob and the level 
will be set as the vertical mid point values of the amplitude of the trigger 
signal. 
Force: Force to create a trigger signal and the function is mainly used in 
"Normal" and "Single" mode. 
Trigger Menu: The button that activates the trigger control menu. 
Trigger Control 
The oscilloscope provides four trigger types: single trigger, alternate trigger, logic 
trigger and bus trigger. Each type of trigger has different sub menus. 
Two ways to enter trigger mode: 