Instructions
4.Advanced User Guidebook 
29 
(4)  Select 20M in the right menu. The bandwidth is limited to 20 MHz. The frequencies 
above 20MHz will be rejected. 
6.  To set the input impedance 
To reduce the circuit load caused by the interaction of the oscilloscope and the circuit to 
be tested, you can set the input impedance. 
Taking the Channel 1 for example, the operation steps are shown as below: 
(1) Push the CH1 button to show CH1 SETUP menu. 
(2) Select Input IMP in the bottom menu, push to select 1MΩ or 50Ω. 
1MΩ: The input impedance of the oscilloscope is rather high and the current flowing 
into the oscilloscope from the circuit under test can be ignored. 
50Ω: Match the  oscilloscope  with  a device with  50  Ω  output  impedance. The max 
input voltage cannot exceed 5 Vrms. 
Use Mathematical Manipulation Function  
The Mathematical Manipulation  function is  used to show the results of the addition, 
multiplication, division  and  subtraction operations  between  two channels, the FFT 
operation  for a channel, advanced math feature including Intg, Diff, Sqrt, user defined 
function, and digital filter. Press the Math button to display the menu on the bottom. 
The Waveform Calculation menu: 
Function Menu 
Setting 
Description 
Dual Wfm 
Math 
Factor1 
CH1 
CH2 
Select the signal source of the factor1 
Sign  + - * /  Select the sign of mathematical manipulation 
Factor2 
CH1 
CH2 
Select the signal source of the factor2 
Vertical 
(div) 
Turn the 
M
 knob to adjust the vertical position of the Math 
waveform 
Vertical 
(V/div) 
Turn the M knob to adjust the vertical division of the Math 
waveform 
FFT 
Source 
CH1 
CH2 
Select CH1 as FFT source. 
Select CH2 as FFT source. 