Figure 3-9 Attenuation Switch   
Caution:  
When the attenuation switch is set to 1X, the probe will limit the bandwidth 
of the oscilloscope in 5MHz. To use the full bandwidth of the oscilloscope, 
the switch must be set to 10X. 
Identify the Probe Attenuation Coefficient Automatically   
The oscilloscope can identify the probe attenuation coefficient of the 100:1 (impedance 
5K±20%)  or  10:1  (impedance  10K±20%)  probe  with the identifying  pin.  When you 
attach  the probe,  the oscilloscope set the attenuation  automatically on the oscilloscope 
vertical menu for the channel to match the probe. 
For example, if you attach a 10:1 probe with the identifying pin, the screen will prompt 
"The probe attenuation factor is X10", and set the attenuation to 10X automatically on the 
oscilloscope vertical menu for the channel. 
How to Use the Probe Safely  
The safety guard ring around the probe body protects your finger against any  electric 
shock, shown as Figure 3-10. 
Figure 3-10 Finger Guard   
Warning: 
To avoid electric shock, always  keep your finger behind the safety guard 
ring of the probe during the operation. 
To  protect you from suffering  from the electric shock, do not touch any 
metal part of the probe tip when it is connected to the power supply. 
Before making any measurements, always 
connect the probe to the 
instrument and connect the ground terminal to the earth. 