2.  Check the displayed waveforms and regulate the probe till a correct compensation is 
achieved (see Figure 3-7 and Figure 3-8).   
Overcompensated    Compensated correctly    Under compensated 
Figure 3-7 Displayed Waveforms of the Probe Compensation   
3.  Repeat the steps mentioned if needed.   
Figure 3-8 Adjust Probe   
How to Set the Probe Attenuation Coefficient  
The probe has several attenuation coefficients, which will influence the vertical scale 
factor of the oscilloscope. 
To change or check the probe attenuation coefficient in the menu of oscilloscope: 
(1)  Push the function menu button of the used channels (CH1 or CH2 button). 
(2)  Select Probe in the bottom menu; select Attenu in the right menu, turn the M knob to 
select the proper value corresponding to the probe. 
This setting will be valid all the time before it is changed again.  
Caution:  
The default attenuation coefficient of the probe on the instrument is preset to 
10X. 
Make sure that the set value of the attenuation switch in the probe is the 
same as the menu selection of the probe attenuation coefficient 
in the 
oscilloscope. 
The set values of the probe switch are 1X and 10X (see Figure 3-9).   