Instructions
3.Junior User Guidebook 
11 
select Default in the bottom menu. The default attenuation coefficient set value of the 
probe in the menu is 10X. 
2.  Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope 
with CH1 Channel.  
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten 
the probe with rotating it to the right side.  
Connect the probe tip and the ground clamp to the connector of the probe 
compensator. 
3.  Push the Autoset Button on the front panel. 
The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in 
several seconds (see Figure 3-6).  
Figure 3-6 Auto set   
Check CH2 by repeating Step 2 and Step 3. 
How to Implement the Probe Compensation  
When connect the probe with any input channel for the first time, make this adjustment to 
match the probe with the input channel. The probe which is not compensated or presents a 
compensation deviation will result in the measuring error or mistake. For adjusting the 
probe compensation, please carry out the following steps: 
1.  Set the attenuation coefficient of the probe in the menu as 10X and that of the switch 
in the probe as 10X (see "How to Set the Probe Attenuation Coefficient" on P12), and 
connect the probe with the CH1 channel. If a probe hook tip is used, ensure that it 
keeps in close touch with the probe. Connect the probe tip with the signal connector 
of the probe compensator and connect the reference wire clamp with the ground wire 
connector of the probe connector, and then push  the  Autoset  button on the front 