Instructions
10.Technical Specifications 
129 
Data Bits 
4 bit to 32 bit 
Data Line Setting 
H, L, X 
CAN Trigger 
Signal Type 
CAN_H, CAN_L, TX, RX 
Trigger Condition 
Start of Frame, Type of Frame, Identifier, 
Data, ID & Data, End of Frame, Missing 
Ack, Bit Stuffing Error 
Baud Rate 
Common, Custom 
Sample Point 
5% to 95% 
Frame Type 
Data, Remote, Error, Overload 
Waveform Generator (optional) 
Performance 
Characteristics 
Instruction 
Max Frequency Output 
25 MHz (Sample 125 MS/s) or 50 MHz (Sample 250 MS/s) 
(optional) 
Channel 
2 
Vertical Resolution 
14 bits 
Amplitude Range 
2 mVpp - 6 Vpp 
Waveform length 
8K 
Standard Waveforms 
Sine, Square, Ramp, and Pulse 
Multimeter (optional) 
Performance 
Characteristics 
Instruction 
Full scale 
reading 
3¾ digits (Max 4000 – count) 
Diode 
0 V - 1.5 V 
Input 
impedance 
10 MΩ 
On/Off 
measurement 
<50(±30)Ω beeping 
Capacitance 
51.2nF - 100uF: ±(3%±3 digit) 
Voltage 
DCV: 400mV, 4V, 40V, 400V, 1000V: ±(1%±1digit)  
Max. input: DC 1000V 
ACV: 400mV, 4V, 40V, 400V: ±(1%±3digit)  750V: ±(1.5%±3digit) 
 Frequency: 40Hz - 400Hz, Max. input: AC 750V (virtual value) 
Current 
DCA: 40mA, 400mA: ±(1.5%±1 digit) 4A, 10A: ±(3%±3digit) 
ACA: 40mA: ±(1.5%±3digit) 400mA: ±(2%±1digit) 
 4A, 10A: ±(3%±3digit) 
Resistance 
400Ω: ±(1%±3digit) 4KΩ~4MΩ: ±(1%±1digit)  
40MΩ: ±(1.5%±3digit) 