Instructions
5.Use the Arbitrary Function Generator (Optional) 
101 
Rhombus  Rhombus signal 
Cardiac  Cardiac signal   
Noise  Noise signal 
Frequency Response Analysis 
The Frequency Response Analysis (FRA) feature controls the built-in waveform generator 
to  sweep a sine wave across a range of frequencies while measuring the input to and 
output from a device under test. At each frequency, gain and phase are measured and 
plotted on a frequency response Bode chart.  When the frequency response analysis 
completes, you can move a marker across the chart to see the measured gain and phase 
values at each frequency point. You can also adjust the chart's scale and offset settings for 
the gain and phase plots. 
Note: If the signal is disturbed seriously, it is recommended to select Average in Acquire 
Mode before running the analysis, the averages can only be set as 4 or 16, and then run the 
analysis.   
Push the Utility button, select Function in the bottom menu, select FRA in the left menu. 
The description of FRA Menu is shown as the follows: 
Function 
Menu  
Setting  Description 
FRA 
FRA 
Check to enable FRA function and show FRA scale 
line and chart. 
Transparent 
When unchecked, FRA information is shown on the 
FRA window.   
When checked, FRA information is shown on the 
waveform display area.   
Marker 
Turn Multipurpose knob to move the marker, view 
measured gain and phase values. 
Sweep Step 
Fine 
When unchecked, the sweep step is 0.1dB. 
When checked, the sweep step is 0.01dB. 
Setup 
Input V 
CH1 
Input V is input from CH1. 
Output V is input from CH2. 
(Note: The menus are not for selecting menu items, only 
for showing information.)
Output V 
CH2 
Min Freq 
Fine 
Turn Multipurpose knob to set frequency sweep min 
value. Check to enable fine-tuning. 
Max Freq 
Fine 
Turn Multipurpose knob to set frequency sweep max 
value. Check to enable fine-tuning. 
Amplitude 
Fine 
Turn Multipurpose knob to set waveform generator 
amplitude. Check to enable fine-tuning. 
Note: The amplitude range is 2 mVpp to 6 Vpp. 
Chart 
Gain Scale 
Adjust scale of gain plot, range from 5dB to 500dB. 