Output menu item in the bottom menu sets the output type of Trig Out(P/F) port on 
Rear Panel. In the bottom menu, select Output. The description of Output menu is 
shown as the follows: 
Device and Print Setup menu items set the print output, refer to "How to Print the 
Screen Image" on page 89. 
●  LAN Set 
Using the LAN port, the oscilloscope can be connected with a computer. 
●  Update 
Use the front-panel USB port to update your instrument firmware using a USB 
memory device. Refer to "How to Update your Instrument Firmware" on page 76. 
●  FRA (Frequency Response Analysis) 
If there is a built-in arbitrary function generator (optional), you can use the frequency 
response analysis. Refer to "Frequency Response Analysis" on page 97. 
How to Update your Instrument Firmware 
Use the front-panel USB port to update your instrument firmware using a USB 
memory device. 
USB memory device requirements: Insert a USB memory device into the USB port 
on the front panel. If the icon   appears on the top right of the screen, the USB 
memory device is installed successfully. If the USB memory device cannot be 
detected, format the USB memory device according to the methods in "USB disk 
Requirements" on P60. 
Caution: Updating your instrument firmware is a sensitive operation, to prevent 
damage to the instrument, do not power off the instrument or remove the USB 
memory device during the update process. 
To update your instrument firmware, do the following: 
1.  Push the Utility button, select Function in the bottom menu, select Configure in 
the left menu, select About in the bottom menu. View the model and the currently 
Function 
Menu 
Setting  Description 
Type 
Trig level 
Output trig signal synchronously 
Pass/fail 
Output High Level when Pass , and Low Level when Fail 
AG 
Output 
CH2  Output of  dual-channel  waveform generator 
(optional) 