Instructions
4.Advanced User Guidebook 
74 
Function Menu 
Description 
Self Cal 
Carry out the self-calibration procedure. 
Default 
Call out the factory settings.  
ProbeCh. 
Check whether probe attenuation is good. 
Do Self Cal (Self-Calibration) 
The self-calibration procedure can improve the accuracy of the oscilloscope under 
the ambient temperature to the greatest extent. If the change of the ambient 
temperature is up to or exceeds 5℃, the self-calibration procedure should be 
executed to obtain the highest level of accuracy.  
Before executing the self-calibration procedure, disconnect all probes or wires from 
the input connector. Push the Utility button, select Function in the bottom menu, 
the function menu will display at the left, select Adjust. If everything is ready, 
select Self Cal in the bottom menu to enter the self-calibration procedure of the 
instrument. 
Probe checking 
To check whether probe attenuation is good. The results contain three circumstances: 
Overflow compensation, Good compensation, Inadequate compensation. According to 
the checking result, users can adjust probe attenuation to the best. Operation steps are 
as follows: 
1.  Connect the probe to CH1, adjust the probe attenuation to the maximum. 
2.  Push the Utility button, select Function in the bottom menu, select Adjust in the 
left menu. 
3.  Select  ProbeCh.  in the bottom menu, tips about probe checking shows on the 
screen. 
4.  Select ProbeCh. again to begin probe checking and the checking result will occur 
after 3s; push any other key to quit. 
●  Pass/Fail  
The Pass/Fail function monitors changes of signals and output pass or fail signals by 
comparing the input signal that is within the pre-defined mask. 
Push the Utility button, select Function in the bottom menu, select Pass/fail in the 
left menu. 
Function Menu 
Setting 
Description 
operate 
Enable 
Operate 
Control enable switch 
Control operate switch 
Output 
Pass 
Fail 
Beep 
Stop 
Info 
Signal tested corresponds with the rule 
Signal tested not correspond with the rule 
Beep when it satisfies the rule  
Stop once satisfying the rule 
Control the display status of info frame 