How to Set the Vertical System  
The VERTICAL CONTROLS includes three menu buttons such as CH1, CH2, CH3, 
CH4 and Math, and two knobs such as Vertical Position, Vertical Scale. 
Setting of CH1 - CH4 
Each channel has an independent vertical menu and each item is set respectively based on 
the channel. 
To turn waveforms on or off (channel, math) 
Pushing the CH1, CH2, CH3, CH4, or Math buttons have the following effect: 
• If the waveform is off, the waveform is turned on and its menu is displayed. 
• If the waveform is on and its menu is not displayed, its menu will be displayed. 
• If the waveform is on and its menu is displayed, the waveform is turned off and its menu 
goes away. 
The description of the Channel Menu is shown as the following list: 
Function 
Menu 
Setting 
Description  
Coupling 
DC 
AC 
GROUND 
Pass both AC and DC components of the input signal. 
Block the DC component of the input signal. 
Disconnect the input signal. 
Inverted 
ON 
OFF 
Display inverted waveform. 
Display original waveform. 
Probe 
Attenu 
0.001X 
to 
1000X 
Step by 1 – 2 – 5. Match this to the probe attenuation 
factor to have an accurate reading of vertical scale. 
MeasCurr 
YES 
NO 
If you are measuring current by probing the voltage 
drop across a resistor, choose YES. 
A/V (mA/V)  
V/A (mV/A)  
Turn  the  M  knob to set the Amps/Volts ratio. The 
range is 100 mA/V - 1 KA/V. 
Amps/Volts ratio = 1/Resistor value  
Volts/Amp ratio is automatically calculated. 
Limit  
Full band 
20M 
Get full bandwidth. 
Limit the channel bandwidth to 20MHz to reduce 
display noise. 
1.  To set channel coupling  
Taking the Channel 1 for example, the measured signal is a square wave signal 
containing the direct current bias. The operation steps are shown as below: 
(1) Push the CH1 button to show the CH1 SETUP menu. 
(2) Select Coupling in the bottom menu. 