Instructions
4.Advanced User Guidebook 
53 
How to Operate the Function Menu  
The function menu control zone includes 8 function menu buttons: Measure, Acquire, 
Utility, Cursor, Autoscale, Save, Display, Help and 3 immediate-execution buttons: 
Autoset, Run/Stop, Single. 
How to Implement Sampling Setup 
Push the Acquire button, Acqu Mode, Length  and Intrpl is shown in the bottom 
menu. 
The description of the Acqu Mode menu is shown as follows: 
Function Menu 
Setting 
Description  
Acqu 
Mode 
Sample 
Normal sampling mode. 
Peak Detect   
Use to capture maximal and minimal 
samples. Finding highest and lowest points 
over adjacent intervals. 
It is used for the 
detection of the jamming burr and the 
possibility of reducing the confusion. 
Average  4, 16, 64, 128 
It is used to reduce the random and don't-care 
noises, with the optional number of averages. 
The description of the Record Length menu is shown as follows: 
Function Menu 
Setting 
Description  
Length 
1000 
Choose the record length 
Note: When four channels are turned on, the 
max record length is 10M;  and  max  20M 
for two channels;  max  40M for one 
channel. 
10K 
100K 
1M 
10M 
20M 
40M 
The description of the Intrpl menu is shown as follows: 
Function Menu 
Setting 
Description  
Intrpl 
Sinx/x 
Use sine(x)/x interpolation 
x 
Use linear interpolation 
Interpolation method is a processing method to connect the sampled  points, using 
some points to calculate the whole appearance of the waveform. Select the appropriate 
interpolation method according to the actual signal. 
Sine(x)/x interpolation: Connect the sampled points with curved lines. 
Linear interpolation: Connect the sampled points with straight lines. This method is 