Instructions
4.Advanced User Guidebook 
48 
ID&Data 
Configure 
(Bottom 
menu)
Format 
Select Standard or Extended. 
ID 
Use  the  M  knob  and  Direction 
key on the front panel to set. 
Byte 
Length 
Set the number of bytes with the 
M knob. The range is 1 to 8. 
Data 
Set the data with the M knob 
and  Direction key  on the front 
panel. 
End 
Trigger on the end frame of the data frame. 
Missing 
Ack 
Trigger on Missing Ack. 
Bit 
Stuffing 
Trigger on Bit Stuffing Error. 
Mode 
Holdoff 
Auto 
Normal 
Single 
Acquire waveform even no trigger occurred 
Acquire waveform when trigger occurred 
When trigger occurs, acquire one waveform then stop 
Bus Decoding   
1.  RS232 Decoding 
To decode RS232 signal: 
(1) Connect the RS232 signal to the Signal Input Channel of the oscilloscope. 
(2) Adjust to the proper time base and voltage division. 
(3) In trigger menu, select Bus trigger, and select bus type as RS232, set parameters 
based on the characteristics of the signal, trigger the signal correctly and obtain 
stable display. Refer to "RS232 Trigger" on page 42. 
(4) Push the Decode  button  on the front panel.  Select  bus type as RS232. set 
parameters based on the characteristics of the signal. When the parameters are set 
correctly, the information carried by the signal will be displayed.  
Tip: If there are repetitive menu items in both trigger menu and decoding menu, you 
can set anyone of them, the other will be changed synchronously. 
Note:  
  Use the Trigger Level knob to adjust the  thresholds of bus trigger and bus 
decoding. 
  When decoding, if "Parity" is not set to "None", and the check bit error is 
detected, two red error marks will be displayed in the corresponding position in 
the waveform. 
RS232 Decoding menu list: 
MENU 
SETTING 
INSTRUCTION 
Bus Type  RS232  Set bus type of decoding as RS232. 