Instructions
4.Advanced User Guidebook 
40 
Edge 
Edge
Start timing when the rising edge of the input signal 
passes through the trigger level. 
Start timing when the falling edge of the input signal 
passes through the trigger level. 
Configure  Idle Time 
Set idle time. Idle time means the minimum time of idle 
clock before searching data that can meet trigger 
conditions. Available range is 30ns-10s, default 100ns. 
Mode 
Holdoff 
Auto 
Normal 
Single 
Holdoff 
Reset 
Acquire waveform even no trigger occurred 
Acquire waveform when trigger occurred 
When trigger occurs, acquire one waveform then stop 
100 ns - 10 s, adjust M knob or click   to set time 
interval before another trigger occur, press     
panel button or click   move cursor to choose 
which digit to be set. 
Set Holdoff time as 100 ns 
8.Nth Edge trigger  
The oscilloscope triggers on the Nth edge that appears on the specified idle time. As 
figure shown below, the oscilloscope should trigger on the second falling edge after 
the specified idle time and the idle time should be set to P1/P2/P3/P4 < Idle Time < M. 
Wherein, M, P1, P2, P3 and P4 are positive or negative pulse width participating in 
the counting. 
In Nth Edge Trigger mode, the trigger setting information is displayed on bottom right 
of the screen, for example,  ,indicates that trigger type is Nth 
Edge, trigger source is CH1, -150V is up level or low level threshold. 