13. Troubleshooting
The multimeter was designed using the latest technology and is safe to use. However, problems and malfunc-
tions may still occur. 
This section explains how to troubleshoot common issues:
  Always observe the safety information in these instructions.
Error Possible cause Possible solution
The multimeter does not 
work.
Isthebatteryat? Check the status. Replace the batteries.
The measured value does 
not change.
Have you selected the wrong meas-
urement mode (AC/DC)?
Check the display (AC/DC) and select 
another mode if necessary.
Did you use the wrong measurement 
sockets? 
Compare the connection with the 
display.
Is the hold function enabled? Disable the hold function.
The multimeter cannot take 
measurements in the 10 A 
range.
Is the fuse in the 10 A input defective? Check the F1 fuse.
The multimeter cannot take 
measurements in the mA/
µA range.
The PTC fuse was triggered, which 
limits the measurement current.
Reduce the measuring current or switch 
to the 10 A measuring range.
Erratic measurement dis-
play in the mA/µA range.
The internal resistance of the 
integrated PTC fuse increases slowly 
during a long measurement.
Limit the measurement time. A pause 
in the measurement allows the fuse to 
cool down.
  Repairs other than those described above should be performed only by an authorised specialist. If 
you have questions about the multimeter, please contact our technical support team.