Alternating current (A/AC)
Range Resolution Accuracy
200.0 µA 0.1 µA
±(1.3% + 7)20.00 mA 0.01 mA
200.0 mA 0.1 mA
10.00 A 0.01 A ±(2.6% + 7)
600 V overload protection
Fuses: µA/mA = 2x 160 mA resettable PTC fuses
10 A = High-performance FF12AH1000V ceramic fuse
≤6A=continuousmeasurement,>6A=max.10secsatintervalsof15minutes
The multimeter may display 3 counts when a measuring input is open
TrueRMSpeak(CrestFactor(CF))≤3CFovertheentirerange
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Resistance
Range Resolution Accuracy
200.0Ω* 0.1Ω ±(1.3% + 3)
2.000kΩ* 0.001kΩ
±(1.0% + 8)20.00kΩ 0.01kΩ
200.0kΩ 0.1kΩ
2.000MΩ 0.001MΩ ±(1.6% + 4)
20.00MΩ 0.01MΩ ±(3.0% + 6)
600 V overload protection
Measuring voltage: Approx. 1.0 V, measuring current approx. 0.7 mA
*Accuracyformeasuringrange≤200ΩwascalculatedafterdeductingleadresistancefromtheRELfunction
Battery test
Range Load resistance Resolution Accuracy
1.5 V Approx.15Ω 0.001 V ±(5% + 3)
9 V Approx.1kΩ 0.01 V
600 V overload protection, resettable PTC fuse