d) Direct current ( )
Select < > mode. [ A DC ] will appear on the display.
The main display indicates the measured current.
e) AC + DC current
Select < AC+DC > mode under < >. [ A ] will appear on the display.
The main display indicates the measured current in form of .
The small display alternates between displaying the measured DC and AC current.
f) Surge/Inrush current
The INRUSH function facilitates the measurement of surge/inrush currents on motors.
1. Select < > mode. [ A AC ] will appear on the display.
2. Long press the V+A/INRUSH button to enter the INRUSH mode. will appear on the display.
The main display indicates the measured surge/inrush current.
The small display indicates the measured stable working current.
Surge/inrushcurrentismeasuredduringtherst100ms(seebelowdiagram)basedonthefollowingcriteria:
Range 60 A: 6 A detection threshold, max. 80 A measurement
Range 600 A: 60 A detection threshold, max. 800 A measurement
0 A
TIME
100 ms Sample Window (60 Hz)
Measured Inrush
Current
Current
Detection
Threshold
CURRENT
Stop
Start
Long press the V+A/INRUSH button to exit surge current measurement mode.