Instructions

60
Direct voltage (V/DC)
Range Resolution Accuracy
6.000 V 0.001 V ±(0.6% + 3)
60.00 V 0.01 V
±(0.9% + 6)
600.0 V 0.1 V
600Voverloadprotection;Impedance:10MΩ
Speciedmeasuringrange:5–100%ofthemeasuringrange
Alternating voltage (V/AC)
Range Resolution Accuracy
6.000 V 0.001 V
±(1.0% + 6)60.00 V 0.01 V
600.0 V 0.1 V
600.0 V "LoZ" 0.1 V ±(2.5% + 6)
600.0 V "Motor" 0.1 V ±(2.0% + 6)
Frequencyrange:40–400Hz;600Voverloadprotection;Impedance:10MΩ(LoZ:300kΩ)
Speciedmeasuringrange:5–100%ofthemeasuringrange
Themultimetermaydisplay≤5countsifameasuringinputisshort-circuited.
After using the LoZ feature, leave the multimeter for 1 minute before using it again.
TrueRMSpeak(CrestFactor(CF))≤3CFto600V
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Criteria for phase detection in "Motor" mode: >80–600 V/AC, 50–80 Hz
Direct current (µA/DC)
Range Resolution Accuracy
2000 µA 1 µA ±(0.9% + 6)
600 V overload protection
Automatic current limitation with integrated PTR components.
Alternating current (µA/AC)
Range Resolution Accuracy
2000 µA 1 µA ±(1.5% + 6)
Frequency range: 40–400 Hz; 600 V overload protection
Automatic current limitation with integrated PTR components.