User manual
 
 
 
The multimeter features a folding current clamp (2) for non-contact current measurement. The 
sensors in the clamp detect the magnetic eld surrounded by the live conductors. Measure-
ment of insulated as well as non-insulated conductors and rails is permissible. Make sure that 
the conductor always passes through the middle of the clamp (observe arrow markings).
  Only ever clamp one cable at a time. If you clamp the supply and return conductors 
(i.e. L and N or positive and negative), the currents will cancel one another out and you 
will not get a measurement. If you clamp more than one outer cable (i.e. L1 and L2 or 
positive conductor 1 and positive conductor 2), the currents will add up.
   A cable selector (1) is located at the tip of the clamp so you 
can conveniently isolate and clamp individual conductors 
from a cable harness.
   In case of low currents, you may coil a conductor around 
one arm of the clamp several times. Then divide the 
measuring result by the number of coils to receive the 
correct current value.