User manual
47
b) AC Current Measuring “A  ”
 Do not exceed the maximum permitted input values. Do not touch any circuits 
or parts of circuits if they may be subject to voltages higher than 33 V/ACrms 
or 70 V/DC! Danger to life!
   The maximum permissible voltage in the measuring circuit against ground 
potential must not exceed 600 V in CAT III.
   Observe the required safety notes, provisions and safety measures for 
intrinsic protection.
The current measurement takes place contact-free via the unfoldable current probe (2). 
The sensors in the current probe record the magnetic eld that is surrounded by conductors 
through which current ows. A measurement is permitted on insulated as well as uninsulated 
conductors and current rails. Ensure that the conductor always runs through the current probe 
centred and that the current probe is always closed during measurement.
   Always hold only one conductor with the current probe. If both the supply and return 
conductors (e.g. L and N or plus and minus) are recorded, the currents will cancel each 
other out and you will not get any measuring result. If several outer conductors are 
recorded (e.g. L1 and L2), the currents will add up.
   At very low currents, the conductor can be coiled 
several times around a shank of the current probe to 
increase the overall measuring current. Then divide 
the measured current value by the number of coils 
around the current probe. You will then receive the 
correct current value.










