User manual

43
d) Replacing the fuse
Observe all safety instructions when replacing the fuse!
Make sure to use replacement fuses of the specied type and nominal current. Use of incorrect or
patched fuses or bypassing the holder is prohibited and can result in a re.
The test inputs “V/ma” (7) and “+10 A” (4) are secured against overload. If measurements can no longer be conducted,
the corresponding fuse is in need or replacement.
Proceed as follows to replace a fuse:
- Remove the test leads from the device and turn off the multimeter.
- Remove the 0 ohm adjustment knob (8) from the multimeter by pulling it upward.
-
Loosen the housing screw at the back and carefully push the battery compartment cover (10) upward to remove.
- Remove both side rubber guards from the housing. To do so, start at the battery compartment and pull the guards
from the housing.
- Carefully pull the plastic pin between the battery compartments backward until the housing case unlocks. Care-
fully remove the housing case.
- Replace the ceramic high-performance fuse in question with a new one of the identical type and nominal current.
F 0.5 A / 500 V quick-acting 6.3 x 32 mm
F 10 A / 500 V quick-acting 6.3 x 32 mm
-
Close the device in reverse order and carefully screw the battery compartment cover back on.