Instructions

30
3. Intended use
Measuring and displaying electrical parameters in the measuring category CAT III up to max. 500 V against ground
potential in compliance with EN 61010-1 and all lower categories. The multimeter may not be used in the measur-
ing category CAT IV.
DC and AC voltage measurements up to max. 500 V
Measuring direct current up to 10 A
Measuring frequency up to 25 kHz
Measuring resistance up to 200 MOhm
Audible continuity test
Diode test
Transistor test
The multimeter is operated with commercially available batteries (2x 1.5 micro batteries (AAA, LR03), 1x 9 V block
((6F22, 4922, NEDA1604 or identical battery)). The multimeter may be operated with the specied battery type only.
Batteries with a cell voltage of 1.2 V may not be used. An automatic power-off function prevents premature battery
drain.
Do not operate the multimeter when the housing or battery compartment is open or if the battery compartment cover
is missing.
Measurements in potentially explosive areas (Ex) or wet rooms or in adverse ambient conditions are not permissible.
Adverse ambient conditions are dened as such: Wet conditions or high humidity, dust or ammable gases, vapors or
solvents as well as thunderstorms or similar conditions such as strong electrostatic elds, etc.
Only use test leads and measuring accessories that meet the multimeter’s specications.
The multimeter may be only operated by persons familiar with the applicable regulations for the measurement and
all possible hazards.
Any use other than that described above will damage the product and may involve other hazards such as short-circuit,
re, electric shock, etc. The product as a whole must not be modied or altered!
Read this user manual completely before rst use and keep it for future reference.
Observe the safety instructions at all times!