Instructions
65
Direct voltage V/DC
Range Resolution Accuracy
600.0 mV* 0.1 mV ±(0.8% + 3)
6.000 V 0.001 V ±(0.6% + 3)
6.000 V AC+DC 0.001 V ±(2.0% + 5)
60.00 V 0.01 V ±(0.8% + 3)
60.00 V AC+DC 0.01 V ±(2.0% + 5)
600.0 V 0.1 V
±(0.8% + 3)
1000 V 1 V
*only available via the measuring function "mV"
Overloadprotection1000V;Impedance:10MΩ(mV:≥1000MΩ)
When the measuring input is short-circuited in the mV area, a display of 5 counts is possible;
For AC+DC function, the frequency range for ACV is 40-400 Hz
Alternating voltage V/AC
Range Resolution Accuracy
600.0 mV* 0.1 mV ±(1.3% + 5)
6.000 V 0.001 V ±(1.0% + 4)
60.00 V 0.01 V
±(1.3% + 4)600.0 V 0.1 V
1000 V 1 V
600.0 V "LoZ" 0.1 V ±(2.6% + 3)
600.0 V "Motor" 0.1 V ±(2.0% + 7)
6.000 V - 1000 V "LPF" depending on measurement range ±(2.0% + 3)**
*only available via the measuring function "mV"
**Additional measuring error with the "LPF" function activated
Frequency response 40 Hz - 5 kHz (when the measured voltage frequency is greater than 1 kHz, the accuracy
needstoadd5Counts);Overloadprotection1000V;Impedance:10MΩ(mV:≥10MΩ)
Speciedmeasurementrange:5-100%ofthemeasuredrange
When the measuring input is short-circuited, a display of 5 counts is possible
After use of the LoZ function, a regeneration time of 1 minute is required
TrueRMScrestfactor(CF)≤3CFto600V,≤1.5CFto1000V
TrueRMS crest factor for non-sine-shaped signals plus tolerance addition:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Criteria for phase recognition in the "Motor" mode: >80 V/AC, 40 - 80 Hz