Instructions

62
Signal level:
≤100kHz:200mV-30Vrms
>100 kHz - <1 MHz: 600 mV - 30 Vrms
≥1MHz-<10MHz:1V-30Vrms
10 MHz: 1,8 V - 30 Vrms
Frequency “Hz” (electrical, subfunction of A and V)
Range Resolution Accuracy
40 - 400 Hz 0.1 Hz Notspecied
Signal level: Sensitivity at least 200 mVrms
Pulse ratio “%”
Range Resolution Accuracy
0.1 – 99.9 % 0.1 % Notspecied
Display of the positive half-wave in %
Frequency range: <10 kHz
Input sensitivity >2 Vpp (10.0% - 95.0%)
Diode test
Test voltage Resolution
approx. 12 V/DC 0.001 V
Overload protection: 1000 V; test current: 3.5 mA typ.
Acoustic continuity tester
Measurement range Resolution
600Ω 0.1Ω
≤10Ωpermanentsound;≥50Ω,thebuzzerwon´tbeep
Overload protection: 1000 V
Test voltage approx. 1 V
Test current <0.25 mA
Do not exceed the maximum permitted input values. Do not touch any circuits or parts of circuits if
they may be subject to voltages higher than 33 V/ACrms or 70 V/DC! Danger to life!