User manual

19
PressV and it will be in DC voltage
measurement .
There are 3 measurement modes: DCVDCVACV
and dBm. PressSELECTto select one of them.
When the power is on, the measurement will be on AUTO state. PressRANGE
to select range manually: 8.0000V/80.000V/600.00V
According with the users requirements, pressREL MAX/MIN and
2nd VIEWto do test and record. All of them will be on slave display.
Insert the red test leads into “VHz” terminal and the black test lead into “COM”
terminal. Contact the test points with the test leads. Please refer to Chart 5. The
measured value and relative symbols will be on master and slave screen at the
same time. Please refer to Table 5:
Table 5
StatePressSELECTto select state
Master
D
ispla
y
Slave Displayswitch between them by
pressing2nd VIEW
DCV DCV
DCV+ACV DCV ACV/Hz
dBm dBm Hz/DCV/ACV/DCV+ACV
In dBm testthe default impedance is 600. PressRANGEto change the load
impedance and the selectable values are as below: 4, 8, 16, 32, 50, 75,
93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900,
1000 and 1200. In dBm measurement mode, the decimal point of dBm is
fixed between the second and third digits and REL, MAX/MINand P-H
are disabled.
5.2 AC Voltage measurement
Warning: The measured voltage must not exceed AC 450V because the high
voltage standard of this DMM is IEC 61010 CAT II 600V
Note: Before tests on low voltage, please connect the test leads in
short-circuit and pressRELto make zero for a better accuracy!
PressV to enter AC voltage function
AC voltage has three states of ACVACVHz
and dBm. PressSELECT to choose one of them.
When the power is on, the measurement is
in AUTO state. PressRANGE to select
suitable range. The selectable ranges are as
below: 8.0000V/80.000V/450.00V
Press (【REL,MAX/MIN and
2nd VIEWto do relative test and
record. All of them can be on slave
Chart 5 DC Voltage Measurements