PTM-100 Operation | 6
B-H86.0.02A.DB214-1.0 17 / 26
Display in water bath 36.6 °C
Display in water bath setpoint T.SL = 37.0 °C
T.OF = display zero point correction – setpoint zero point
T.OF = -0.2 °C – 0.0 °C = -0.2 °C
T.SL = (setpoint gradient correction / (display gradient correction – T.OF) – 1) *100
T.SL = (37.0 °C / (36.6 °C – (-0.2)) -1) *100 = 0.54
Outcome of an action
The changed value is saved and the Configuration menu is closed.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.