USER’S MANUAL PSO2020 V1.0.0 www.hantek.
Content General Safety Summary .............................................................................. 1 Chapter 1 Getting Start.................................................................................. 2 1.1 System Requirement ......................................................................................... 3 1.2 Install Software ................................................................................................. 4 1.3 Install Driver .....................................
PSO2020 General Safety Summary Review the following safety precautions carefully before operate the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards use the device as specified by this user’s guide only. To Avoid Fire or Personal Injury Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use. Connect and Disconnect Properly.
PSO2020 Chapter 1 Getting Start The oscilloscope is small, lightweight, no external power required, portable oscilloscopes! The oscilloscopes is ideal for production test, research and design and all of the applications involving analog circuits test and troubleshooting, as well as education and training.
PSO2020 1.1 System Requirement To run the oscilloscope software, the needs of computer configuration are as follows: Minimum System Requirements Operating System Window XP /7/8/10/Vista 32bits and 64bits Processor Upwards of 1.00G processor Memory 256M byte Disk Space 500M disk free space Screen resolution 800 x 600 Recommended Configuration Windows XP SP2 System 2.
PSO2020 1.2 Install Software Caution: You must install the software before using the oscilloscope. 1. While in Windows, insert the installation CD into the CD-ROM drive. 2. The installation should start up automatically. Otherwise in Windows Explorer, switch to the CD-ROM drive and run Setup.exe. 3. The software Installation is started. Click 'Next' to continue.
PSO2020 4. Choose a destination directory. Click 'Next' to continue. 5. Check the setup information. Click Next to start copying of files.
PSO2020 6. This Status dialog is displayed during copying of files. 7. Updating Your System Configuration. 8. The installation is complete.
PSO2020 1.3 Install Driver 1. Connect the A-Type Plug of USB cable to your PC’S USB port. 2. New hardware is found. 3. New hardware search wizard starts. 5. New hardware search wizard starts to search the driver.
PSO2020 6. New hardware wizard installs “DRIVER”.
PSO2020 7. The wizard has finished installing for “DRIVER”. 8. Your new hardware is installed and ready to use.
PSO2020 1.
PSO2020 1.5 General Check Please check the instrument as following steps after receiving an oscilloscope: Check the shipping container for damage: Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. Check the accessories: Accessories supplied with the instrument are listed in "Accessories" in this guide.
PSO2020 1. Attach the tip of probe and ground nip to the Connector of Probe. A square wave will be displayed within a several seconds. (Approximately 1 kHz, 2V, peak- topeak). 1.7 Self Calibration The self calibration routine lets you optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but you should always run the routine if the ambient temperature changes by 5℃ or more.
PSO2020 Chapter 2 Operating Basics The User’s Interface The Menu System The Vertical System The Horizontal System The Trigger System Input Connector User’s Manual 13
PSO2020 2.1 The User’s Interface Click the software icon on the desk after you finished the software setting and equipment connecting. Then a user interface will be showed as follows: In addition to displaying waveforms, the display area is filled with many details about the waveform and the oscilloscope control settings. 1. The Main Menu All settings can be found in the main menu. 2. The Toolbar 3. Control Panel 4. DVM 5. The Horizontal Panel The user can change Time/Div, format in the panel.
PSO2020 12. Output Window 13. The markers show the reference points of the displayed waveforms. If there is no marker, the channel is not displayed. 14. Measure Window. 15. Trige state 16. "Running":the program found the hardware and the hardware works normally. "Searching Device":not find the device and the program is searching the device. 2.
PSO2020 Keep clicking:AUTO SET “+” button 4、TIME (T ) “-” button R/S button 5、combination button 6、- “+” & “R/S” “-” Single click:Time increase Keep clicking:continuous increase with Time Single click:Time decrease Keep clicking:continuous decrease with Time Single click:RUN or STOP Keep clicking:AUTO SET Single click:LED lamp switch - 2.3 The Menu System The Main Menu 1. File: Load or Save data, setup 2.
PSO2020 3. Horizontal: Set the horizontal system 4. Vertical: Set the vertical system 5. Trigger: Set the trigger system 6. Display: The waveform display setup 7. Cursor: Cursor measure setup 8. Utility: Utility setting 9.
PSO2020 2.4 The Vertical System Click “Vertical” in main menu The following figure shows the vertical system menu. It shows the vertical parameters setting. 1. CH1: Leads to the setting of the CH1 parameters Use the left mouse button to dragging the channel icon to set the vertical position offset precisely. 2.5 The Horizontal System Click “Horizontal” in main menu The following figure shows the Horizontal System window. It shows the horizontal parameters settings. 1.
PSO2020 2.6 The Trigger System Click “Trigger” in main menu. Mode: Sets the trigger mode to Edge trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE Source: Selects the trigger source to CH1. Slope: Selects the edge trigger slope to Positive or Negative slope The user can drag the trigger icon to adjust the trigger level and the trigger level will be displayed on the top right corner. 2.7 Input Connectors Input Connectors: CH 1: Input connectors for waveform display. GND.
PSO2020 Chapter 3 Understanding Function Set Oscilloscope Set Vertical System Set Horizontal System Set Trigger System Measure Signal Utility Function The Display System Save/Load Print User’s Manual 20
PSO2020 3.1 Set Oscilloscope Save Setup The oscilloscope software saves the current setup before you close the oscilloscope software. The oscilloscope recalls this setup the next time you run the software. You can use the “Save Setup” menu to permanently save up to several different setups. Load Setup The oscilloscope can recall the last setup before the oscilloscope software was running, any saved setups, or the factory setup. You can use the “Load Setup” menu to permanently recall a setup.
PSO2020 You can left click and drag the mouse on the green knob or the yellow knob to change the voltage. Invert The invert function turns the displayed waveform 180 degrees, with respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted. Click “Invert” in vertical.
PSO2020 3.2.1 Set FFT Fast Fourier Transform Function To use the FFT mode, you need to click in Toolbar to open the FFT setup window. The FFT setup window: You can select the Source channel, Window algorithm, FFT number, and FFT Zoom factor. It displays only one FFT spectrum at a time.
PSO2020 Selects the channel used as the FFT source Window: Selects the FFT window type Scale: Selects the vertical scale units Vert Scale: Selects the vertical scale HORI Scale: Selects the FFT zoom factor Sets horizontal scale, If the information is more than the quantity displayed to the display window, more information is displayed by using scroll bar. You can click the “ON/OFF” button to start FFT math, also you can click this button to stop FFT math. 3.2.
PSO2020 The Reference Waveform Display Window: Note: If you turn on the “Reference” channel, the load file window will appear. 3.3 Set Horizontal System The following figure shows the Horizontal System window. It shows the horizontal parameters settings. 1. Time/DIV: leads the setting of the time base parameters 2. Format: leads the setting of the horizontal format parameters The time base ranges of the oscilloscope is listed as follows.
PSO2020 from 4ns/div to 1hour/div. In the “Format” menu of the “Horizontal” menu, select the display mode to Y-T. 3.4 Set Trigger System Click “Trigger” in main menu. The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms. If the oscilloscope wants to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point.
PSO2020 Single: Acquire waveform when trigger occurred then stop Slope: Set the slope to Rising (+) or Falling (-). Rising: Trigger on rising edge Falling: Trigger on falling edge 3.5 Measure Signal 3.5.1 Cursor Click “Cursor” in main menu. This method allows you to take measurements by moving the cursors 1. Source The user can set the source to CH1. 2. Type There are four types of cursors: Cross, Vertical and Horizontal.
PSO2020 2) Vertical The Vertical cursors appear as vertical lines on the display and measure the vertical parameters. The Vertical cursor display window: 3) Horizontal The Horizontal cursors appear as horizontal lines on the display and measure the horizontal parameters.
PSO2020 3.5.2 Measure Click “Measure” in main menu. The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time measurements). 1.
PSO2020 Middle: Voltage of the 50% level from base to top RMS: The Root Mean Square voltage over the entire waveform Amplitude: Amp = Base – Top, Measured over the entire waveform Mean: The arithmetic mean over the entire waveform Cycle Mean: The arithmetic mean over the first cycle in the waveform Preshoot: Positive Overshoot = (Max - Top)/Amp x 100 %, Measured over the entire waveform Overshoot: Negative Overshoot = (Base - Min)/Amp x 100 %, Measured over the entire waveform 2.
PSO2020 Note: The results of the automatic measurements will be displayed on the bottom of the screen. Maximum 8 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left, out of screen. 3.6 The Utility Function Click the “Menu -> Utility” to get into the “Utility” menu.
PSO2020 Interpolation Language 3.6.1 Default Setup Click “Default Setup” in “Utility” menu to load default setups When you click the Default Setup in Utility menu, the oscilloscope displays the CH1 waveforms and removes all other waveforms. The oscilloscope set up for normal operation when it is shipped from the factory and can be recalled at anytime by user. The Factory Setup function does not reset the following settings: Language option Date and time 3.6.
PSO2020 Click “Utility->Interpolation”.
PSO2020 The Sin(x)/x Interpolation: Note: The default interpolation mode is Sin(x)/x.
PSO2020 3.7 The Display System Click “Display” in main menu. The following figure shows the display system menu. Displays parameters setting. Type Grid Intensity 3.7.1 Type Click “Type” in “Display” menu. If the Vectors type mode is selected, the waveform will be displayed as following figure. If the Dots type mode is selected, the waveform will be displayed as following figure.
PSO2020 3.7.2 Grid Click “Display” in main menu.
PSO2020 Without the grid show: 3.7.
PSO2020 Click “Display->Intensity” in main menu The following figure shows the intensity dialog. It shows the display parameters setting. You can change the grid and waveform color intensity in this dialog.
PSO2020 3.8 Save/Load Menu Click “File” in main menu to save waveform, setups and screen 1. Save Data Save waveform data 2. Save Setups Save the current oscilloscope setups to file Load Menu Click “File” in main menu to recall saved waveform, setup 1. Load Data Load the waveform 2.
PSO2020 3.9 Print Click “Print” in File. 1. Click “Print” in “File” menu to set the printer to print the current waveform. 2. Click the “PrintPreview” in “File” menu to get into the Preview window. In “PrintPreview” window, use the button to change the size of the waveform graph. Click the “Close” button to turn this window off and click the “Print” button to print the report.
PSO2020 Chapter 4 Application Example Simple Measurement Capturing a Single-Shot Signal Taking Cursor Measurement User’s Manual 41
PSO2020 4.1 Simple Measurement To acquire and display a signal, please do the steps as follows: Connect signal to CH1 by using probe The oscilloscope set the vertical, horizontal, and triggers controls at the best status automatically. Also, you can adjust the controls to meet your measurement to optimize the waveform display. 1. To measure the frequency and “Vpp”, you can do these steps as follows: 2. Click the button, the frequency of the signal display on the bottom of the waveform interface. 3.
PSO2020 The results of measure display in Information output. 4.2 Capturing a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowledge of the signal in order to set up the trigger level and slope correctly. For example, if the event is derived from 3.3V COMS logic, a trigger level of 1.2 or higher Volts should work on a rising edge. Do these steps as follows: 1. Set up the trigger in the Trigger Menu, or in the Trigger Setting window. 1) Adjust the Trigger Mode to Edge.
PSO2020 6) When the trigger conditions are met, data appears on the display representing the data points that the oscilloscope obtained with one acquisition. 4.3 Taking Cursor Measurements Use cursors to make time and amplitude measurements on a waveform quickly. Measure the Peak Frequency or Time of the First Sine Waveform Do these steps: 1. Click “Cursor->Source”, select CH1. 2. Click “Cursor->Type”, select Vertical. 3. Push left mouse button, and the vertical lines appear. 4.
PSO2020 5. Release the left mouse button, the voltage difference will be shown at the status bar. Note: Click “Cursor->Type”, select “Cross”, you can measure time and amplitude at one time.
PSO2020 Appendix Appendix A: Specifications Appendix B: General Maintenance User’s Manual 46
PSO2020 Appendix A: Specification Specifications Table: Input Max. sample rate Real-time sampling:96MS/s Channels Bandwidth 1 Channels 20MHz (-3dB) Vertical resolution 8 bits/channel Gain range 20mV~20V/div Range 8 divisions Offset level +/-4 divisions Coupling DC/AC/GND Offset increments 0.02 div Impedance 1M ohm DC accuracy +/-3% Max. Input 50V Max.
PSO2020 Interface Universal Serial Bus(USB 2.0) Power No external power source required.
PSO2020 Appendix B: General Maintenance General Care Do not store or leave the oscilloscope where the device will be exposed to direct sunlight for long periods of time. Caution To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents. To avoid damages to the surface of the device or probes not use any abrasive or chemical cleaning agents. Cleaning Inspect the device and probes as often as operating conditions require.