User manual

Specifications
MSO-5000B Digital Storage Oscilloscope User Manual 73
Maximum Input
Voltage
Overvoltage Category
Maximum Voltage
CAT I and CAT II
300V
RMS
(10×), Installation Category
CAT III
150V
RMS
(1×)
Installation Category II: derate at 20dB/decade above 100kHz to 13V peak
AC at 3MHz* and above. For non-sinusoidal waveforms, peak value must
be less than 450V. Excursion above 300V should be of less than 100ms
duration. RMS signal level including all DC components removed through
AC coupling must be limited to 300V. If these values are exceeded, damage
to the oscilloscope may occur.
Measurements
Cursors
Voltage difference between cursors: V
Time difference between cursors: T
Reciprocal of T in Hertz (1/ΔT)
Automatic
Measurements
Frequency, Period, Mean, Peak-to-peak, Cycle RMS, Minimum, Maximum,
Rise Time, Fall Time, Positive Width, Negative Width
General Specifications
Display
Display Type
7 inch 64K color TFT (diagonal liquid crystal)
Display Resolution
800 horizontal by 480 vertical pixels
Display Contrast
Adjustable (16 gears) with the progress bar
Probe Compensator Output
Output Voltage,
typical
About 5Vpp into ≥1MΩ load
Frequency, typical
1kHz
Power Supply
Supply Voltage
100-120VAC
RMS
10%), 45Hz to 440Hz, CAT
120-240VAC
RMS
10%), 45Hz to 66Hz, CAT
Power Consumption
<30W
Fuse
2A, T rating, 250V
Environmental
Temperature
Operating: 32 to 122 (0 to 50)
Nonoperating: -40 to 159.8 (-40 to +71)
Cooling Method
Convection
Humidity
+104 or below (+40 or below): ≤90% relative humidity
106 to 122 (+41 to 50): 60% relative humidity
Altitude
Operating and Nonoperating
3,000m (10,000 feet)
Random Vibration
0.31g
RMS
from 50Hz to 500Hz, 10
minutes on each axis
Nonoperating
2.46g
RMS
from 5Hz to 500Hz, 10
minutes on each axis
Mechanical Shock
Operating
50g, 11ms, half sine