43
h) Capacitance
Range Accuracy Resolution
40.00 nF Under REL status:
±(4 % reading +10 digits)
10 pF
400.0 nF Under REL status:
±(4 % reading +3 digits)
100 pF
4.000 µF 1 nF
40.00 µF ±(4 % reading +3 digits) 10 nF
100.0 µF ±(5 % reading +10 digits) 100 nF
Overload protection: 600 V
i) Frequency / Duty Ratio
Range Accuracy Resolution
10 Hz – 10 MHz ±(0.7 % reading +4 digits) 0.001 Hz – 0.001 MHz
0.1 – 99.9 % NA 0.1 %
Overload protection: 600 V
Input amplitude a: (DC level is zero)
300 mV ≤ a ≤ 30 Vrms (for 10 Hz –1MHz)
600 mV ≤ a ≤30 Vrms (for >1MHz – 10MHz)