nel C. For this purpose, set the switch to "HIGH”. If the measured
signals have a high amplitude (88 Vrms = 250 Vpp max.) set the
switch to "LOW”. In doing so, the measurement input is not overrun
(= overdriven).
B4 Adjusting the sampling time and resolution (last digit)
By applying the switch "GATE TIME” and the push-button "RESO”
you can change the sampling rate and the resolution for the three
channels "A”, "B” and "C”. This is done as follows:
Set the switch "GATE TIME” from "FAST” to "SLOW”. Immediately
afterwards, press the push-button "RESO” (= resolution). Subse-
quently, "—SEL.1—" appears on the display. If you apply the push-
button "RESO” again, "—SEL.2—" is displayed. In this context,
observe the following table:
Range Switch position Resolution Sampling time
GATE TIME
2500 MHz FAST 1000 Hz 0,50 s
(CH A) SLOW 100 Hz 2,75 s
SLOW (SEL.1) 200 Hz 1,50 s
SLOW (SEL.2) 500 Hz 0,75 s500 MHz
FAST 100 Hz 0,75 s
(CH B) SLOW 10 Hz 6,00 s
SLOW (SEL.1) 20 Hz 5,00 s
SLOW (SEL.2) 50 Hz 1,50 s
10 MHz FAST 10 Hz 0,50 s
(CH C) SLOW 1 Hz 1,25 s
SLOW (SEL.1) 0,2 Hz 6,00 s
SLOW (SEL.2) 0,1 Hz 11,00 s