User manual
21
CAT I Measurement Category I: For measuring circuits of electrical and electronic equipment that is not directly
supplied with a mains voltage (e.g. battery-operated devices, protective extra-low voltage, signal and con-
trol voltages).
CAT II Measurement Category II: For measuring electrical and electronic devices that are directly supplied with a
mains voltage via a mains plug. This category also includes all lower categories (e.g. CAT I for measuring
signal and control voltages).
CAT III Measurement Category III: For measuring circuits of installations in buildings (e.g. electric sockets or sub-
distributions). This category also includes all lower categories (e.g. CAT II for measuring electrical devices).
Measuring in CAT III is only permitted with test prods with a maximum exposed contact length of 4 mm or
with cover caps over the test prods.
CAT IV Measurement Category IV: For measuring at the origin of a low-voltage installation (e.g. mains distribution,
electricity provider’s transfer points to homes) and outdoors (e.g. when conducting tasks on underground
cables or overhead lines). This category also includes all smaller categories. Measuring in CAT IV is only
permitted with test prods with a maximum exposed contact length of 4 mm or with cover caps over the test
prods.
Earth potential
3. Intended use
• This digital oscilloscope visualises electrical parameters and signals.
• There are two independent input channels.
• The test lead is connected to the oscilloscope via two BNC sockets.
• 20.3 cm (8”) colour TFT display (800 x 600 pixels)
• Built-in help system
• Measures and displays electrical parameters in overvoltage category CAT II (up to 300 V, DC and AC test prod) with
an input impedance of 1 M Ohm at earth potential. Complies with the EN 61010-1 standard and all lower categories.
• 100 MHz bandwidth, 1 GS / s sample rate
• 40 M record length
• Waveform refresh rate: 75000 Wfms / s
• Multiple trigger types
• Built-in digital multimeter:
- Measures and displays electrical parameters in measurement category CAT II (up to 600 V) and CAT 1 (up to
1000 V). Complies with the EN 61010-1 standard and all lower categories.
- The highest range: Measures DC voltages up to 400 V
- The highest range: Measures alternating voltages up to 400 V at a frequency of 40 to 400 Hz
- Measures direct and alternating currents up to 10 A