User manual

11.Technical Specifications
125
Performance Characteristics
Instruction
Scanning speed (S/div)
2ns/div - 1000s/div,
step by 1 2 - 5
Sampling rate / relay
time accuracy
±1 ppm
Interval(T) accuracy
(DC - 100MHz)
Single
±(1 interval time+1 ppm×reading+0.6 ns);
Average>16
±
(1 interval time +1 ppm×reading+0.4 ns)
Vertical system
Sensitivity
1 mV/div10 V/div
Displacement
±2 V (1 mV/div 50 mV/div)
±20 V (100 mV/div 1 V/div)
±200 V (2 V/div 10 V/div)
Analog bandwidth 100 MHz200 MHz300 MHz
Single bandwidth
Full bandwidth
Low Frequency
10 Hz (at input, AC coupling, -3 dB)
Rise time (at input,
Typical)
≤ 3.5 ns
DC gain accuracy
1 mV
3%
≥ 2 mV
2%
DC accuracy (average)
Delta Volts between any two averages of
≥16 waveforms acquired with the same
scope setup and ambient conditions (V):
±(3% rdg + 0.05 div)
Waveform inverted ON/OFF
Measurement
Cursor
V,
T,
T&
V between cursors,
auto cursor
Automatic
Period, Frequency, Mean, PK-PK, RMS,
Max, Min, Top, Base, Amplitude, Overshoot,
Preshoot, Rise Time, Fall Time, +Pulse
Width, -Pulse Width, +Duty Cycle, -
Duty
Cycle, Delay A→B , Delay A→B
, Cycle
RMS, Cursor RMS, Screen Duty, Phase
,
+Pulse Count, -Pulse Count, Rise
Edge
Count, Fall Edge Count
, Area, and Cycle
Area.
Waveform Math , , *, / ,FFT
Waveform storage 100 waveforms
Lissajous
figure
Bandwidth
Full bandwidth
Phase
difference
±3 degrees