User manual
Digital Storage Oscilloscope 53
1. Plug the USB flash disc in which a firmware program is saved in USB Host interface
on the front panel of the oscilloscope.
2. Press [Utility] button to skip to the Utility menu.
3. Press Update->Update Firmware softkey.
4. Select the file and press Universal Knob to confirm. Then press Start updating
softkey to update the firmware.
Reboot the machine after finishing upgrading, and the software version is upgraded.
The oscilloscope should be self-corrected once after upgrading.
2.12.2
Self Calibration
The self calibration routine helps optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but should always run it if
the ambient temperature changes by 5℃ or more. For a more accurate calibration,
please power on the oscilloscope and wait for 20 minutes until it has adequately
warmed up.
To compensate the signal path, disconnect any probes or cables from the front-panel
input connectors. Then, push the [Utility] button, select Calibrate option and follow the
directions on the screen.
2.12.3
Pass/Fail
“Pass/Fail” is used for judging whether the input signal is in a built rule range and
outputting the past or failed waveform so as to detect the change condition of the
signal.
Page 1 of the pass/fail function menu:
Option
Setup
Description
Pass/Fail
ON
OFF
Run the pass/fail function.
Stop the pass/fail function.
Source CH1~CH4 Select the signal input channel.
Regular
Skip to a rule setup menu.
Message
OFF
ON
Close waveform pass/fail time display information.
Open waveform pass/fail time display information.
Stop
On
Off
Enter STOP state if output exist.
Continue to run if output exists.
Page 1/2
Skip to page 2 to the pass/fail menu.