Basic Operation
DSO-1000D Digital Oscilloscope User Manual 39
Next Page
Grid
Dotted line
Real line
OFF
Off only displays the horizontal and vertical coordinates
at the center graticule on the screen.
Grid Intensity
0-15 16 ranks adjustable, with a progress bar to
display.
5.6 Fast Action Buttons
AUTOSET: Automatically set the oscilloscope controls to generate a usable display of the input
signals. Refer to the following table for relative content.
SINGLE SEQ: Acquire a single waveform and then stop the acquisition.
RUN/STOP: Continuously acquire waveforms or stop the acquisition.
HELP: Display the Help menu.
DEFAULT SETUP: Recall the factory setup. Refer to the following table for relative content.
Save to USB Disk: Save all displays on the screen to a USB device, just like the screen capturing
function of a computer.
5.6.1 AUTOSET
Autoset is one of the advantages digital oscilloscopes have. When you push the AUTOSET button,
the oscilloscope will identify the type of waveform (sine or square wave) and adjust controls
according to input signals so that it can accurately display the waveform of the input signal.
Functions Settings
Acquire Mode Adjusted to Normal or Peak Detect
Cursor Off
Display Format Set to YT
Display Type Set to Vectors for an FFT spectrum; otherwise, unchanged
Horizontal Position Adjusted
SEC/DIV Adjusted
Trigger Coupling Adjusted to DC, Noise Reject or HF Reject
Trigger Holdoff Minimum
Trigger Level Set to 50%
Trigger Mode Auto
Trigger Source Adjusted; Autoset can not be used for the EXT TRIG signal
Trigger Slope Adjusted
Trigger Type Edge
Trigger Video Sync Adjusted