Instructions
37
Return : The query returns 5,6,7,8
Example :TRIGger:UART:WIDTh 5 /*set data width to 5*/
:TRIGger:UART:WIDTh? /* The query returns 5*/
TRIGger:UART:STOP
Syntax :TRIGger:UART:STOP <stop>
:TRIGger:UART:STOP?
Description :Set or query the stop bit when the trigger condition is
ERRor in UART trigger
Parameter :<stop> ::= <1|2>
Return : The query returns 1 or 2
Example :TRIGger:UART:STOP 2 /*set stop bit to 2*/
:TRIGger:UART:STOP? /* The query returns 2*/
TRIGger:UART:PARIty
Syntax :TRIGger:UART:PARIty <parity>
:TRIGger:UART:PARIty?
Description : Set or query the parity type when the trigger condition
is ERRor or PARity in UART trigger.
Parameter :<parity> ::= < NONE| ODD| EVEN>
Return : The query returns NONE| ODD| EVEN
Example :TRIGger:UART:PARIty EVEN /*set parity is EVEN*/
:TRIGger:UART:PARIty? /* The query returns EVEN */
TRIGger:UART:WHEN
Syntax :TRIGger:UART:WHEN <when>
:TRIGger:UART:WHEN?
Description : Set or query the trigger condition in UART trigger
Parameter :<when> ::= < EQUAl|NEQUal|GREAt|LESS>
Return : The query returns EQUAl,NEQUal,GREAt,LESS
Example :TRIGger:UART:WHEN LESS /*set trigger condition LESS */
:TRIGger:UART:WHEN? /* The query returns LESS */
TRIGger:UART:IDLe
Syntax :TRIGger:UART:IDLe <when>
:TRIGger:UART:IDLe?
Description : Set or query the idle level in UART trigger
Parameter :<when> ::= < HIGH| LOW>
Return : The query returns HIGH,LOW
Example :TRIGger:UART:IDLe HIGH /*set idle level HIGH*/