Instructions
17
TRIGger:HOLDoff
Syntax :TRIGger:HOLDoff <value>
:TRIGger:HOLDoff?
Description : Set or query the trigger holdoff time
Parameter :<value> ::= holdoff time(unit is s)
Explanation : Trigger holdoff can be used to stably trigger the complex
waveforms (such as pulse series). Holdoff time is the time
that the oscilloscope waits before re-arming the trigger
circuitry. The oscilloscope will not trigger until the
holdoff time expires.
Return : The query returns current average number of points
Example :TRIGger:HOLDoff 0.0000002 /*Set the trigger holdoff time
to 200ns*/
:TRIGger:HOLDoff? /* The query returns 2.000e-07*/