Digital Storage Oscilloscope 55
Auto Scale: Automatically set the oscilloscope controls to generate a usable display of
the input signals. Refer to the following table for relative content.
Single: Acquire a single waveform and then stop the acquisition.
Run/Stop: Continuously acquire waveforms or stop the acquisition.
Help: Display the Help menu.
Default: Recall the factory setup. Refer to the following table for relative content.
2.13.1 AUTO SCALE
Auto Scale is one of the advantages digital oscilloscopes have. When you push the
Auto Scale button, the oscilloscope will identify the type of waveform (sine or square
wave) and adjust controls according to input signals so that it can accurately display
the waveform of the input signal.
Functions Settings
Acquire Mode
Adjusted to Normal or Peak Detect
Cursor Off
Display Format Set to YT
Display Type Set to Vectors for an FFT spectrum; otherwise, unchanged
Horizontal Position
Adjusted
SEC/DIV
Adjusted
Trigger Coupling
Adjusted to DC, Noise Reject or HF Reject
Trigger Holdoff
Minimum
Trigger Level
Set to 50%
Trigger Mode
Auto
Trigger Source
Adjusted; Auto Scale can not be used for the EXT TRIG
signal
Trigger Slope
Adjusted
Trigger Type
Edge
Trigger Video Sync
Adjusted
Trigger Video
Standard
Adjusted
Vertical Bandwidth
Full
Vertical Coupling
DC (if GND was chosen before); AC for the video signal;
otherwise, unchanged
VOLTS/DIV
Adjusted