User manual

DSO-1052 USB
DSO-1082
USB
Over comp
ensated
Under Compensated
If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe
for the fattest square wave being displayed on the oscilloscope.
Repe
at if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the
insulated cable, and do not touch the metallic portions of the probe head while it is
connected with a voltage source.
1.6 Self Calibration
The self calibration routine lets you
optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run
the routine if the ambient temperature changes by 5v or more. For accurate calibration,
power on the oscilloscope and wait twenty minutes to ensure it is warmed up. To
compensate the signal path, disconnect any probes or cables from the input connectors.
Then, access the “Utility -> Calibration” option and follow the directions on the screen.
The self calibration routine takes about several minutes.
US
ERS MANUAL 15