User manual



The voltage and current can be set with an external voltage source or an external, adjustable resistance (abbre-
viation “Poti“) via the installed remote control connection  (10). The remote control is connected
to the rear remote control connection  (10). A plug for the remote connection is enclosed for the
connection.



1. Loosen the side screws of the remote connection socket and remove the front, black contact socket with a small
turning movement.
2. Insert ve connection lines with a cable cross section of at least 0.34 mm
2
from behind through the metal
sleeves. Carefully solder these cables to the solder tails no. 1, 2, 3, 4 and 5 of the black contact sockets. Please
ensure there are no short circuits.
The digits of the soldering lugs are specied on the black insulator.
Mark the loose cable ends with the corresponding contact digits (1-5) to rule out confusion.
Place the black contact sockets in reverse order into the metal sleeves and carefully screw these together.
The pin connections read as follows:
Contact 1 Internal control voltage + 5 V/DC (<50 mA)
Contact 2 Voltage setting
Contact 3 Current setting
Contact 4 Ground potential („Ground“)
Contact 5 On/off output
Contact 6 - 8 Not allocated