(3) Connect CH1 10:1 probe to input terminal, and connect with “CAL”.
(4) Follow Chapter 3 to operate relative components, get waveform on the screen as
figure 4.1
(5) Observe whether the waveform compensation is good, if not , you can adjust the LF
compensation components as shown in figure 4-2
Good Compensate Over Compensate Under Compensate
Fig 4-1
LF COMP
Fig 4-2
4.2 MEASUREMENT
4.2.1 P-P Voltage Measurements
Step :
(1) Input signal to INPUT [18] terminal.
(2) Setup VOLTS/DIV and observe waveform, let waveform display on the screen
within 5 divisions, and turn VAR clockwise to the calibration position.
(3) Adjust level to make waveform steady.
(4) Adjust sweep controls to show at least one cycle waveform on the screen.
(5) Adjust vertical position to make the bottom of waveform lies on a horizontal axis on
the screen. Fig 4-3A.
(6) Adjust horizontal position to make the top of waveform lies center of vertical axis.
Fig 4-3B.
(7) Read the divisions between A-B on vertical direction.
(8) Calculate the signal Vp-p using the formula below :
Vp-p= DIV of vertical direction× Sensitivity
For example, In Fig 4-3, vertical divisions of A-B is 4.1 DIV, sensitivity of the 10:1