3.3.3 VERTICAL SYSTEM OPERATION
(1) VOLTS/DIV switch should be turned to the correct position following the input signal
range. Adjust position [16] to show the whole waveform within the available area.
Adjust with VAR [17] if necessarily, trimming ratio is 2.5:1.
(2) Input coupling options: DCis used for observing a signal with a direct current
content such as logic or static signals, “DC” must be used with low frequency. AC
is used for observing the AC component of signals. (ground) is used to establish
a trace at a zero volt reference.
(3) X-Y OPERATION: When [11] set at EXT/X, The oscilloscope is used for X-Y
operation, at this moment Input [19] is Y-axis with the same sensitivity, Input [13] as
X-axis, [14] can be adjusted continuously within 0.2V/DIV~0.5V/DIV.
3.3.4 TRIGGER SOURCE
In Fig 3-1, [12] provides 3 sources for selection, INT trigger, EXT trigger input from
[13], LINE input from power source.
3.3.5 HORIZONTAL SYSTEM OPERATION
(1) Sweep setup : turn the sweep switch to the correct position according to the signal
frequency and adjust POSITION [7] to show the whole waveform within the available
area. Adjust with VAR [8] if necessarily, trimming ratio is 2.5:1.
(2) There are 3 kinds of Trigger sources: [11] AUTOauto sweep, when a triggering
signal is applied, Level [9] will adjust to correct position. The screen showing a steady