Operation Manual

20
Measuring self-illuminating objects
Self-illuminating objects are illuminated by a separate, „cold“ light source. These objects are e.g. CRT tubes,
LCD/TFT monitors, LED illumination, etc. Hot light sources like common light bulbs and halogen spotlights
may not be measured.
The measuring device is mainly intended for non-self illuminating objects. Due to the
different light sources, the measuring values are not specified in this mode. With self-
illuminating objects, the measuring device can only be applied for comparative measu-
rement.
Proceed as follows for the measurement:
• • Connect the sensor to the measuring device and turn the device on via the „POWER“ button.
The measuring device signals this with a beep. Wait until the „Mesu“ icon goes off on the display. After
activation, the measuring device is always in RGB mode with activated sensor illumination.
Switch the illumination function off by pressing the button „LIGHT ON/OFF“ once. A beep confirms your
selection.
• Press the measuring button (4 or 16) and check the inactivated sensor illumination. If this is lit, turn the illu-
mination in the sensor head off via the button „LIGHT ON/OFF“ (9).
• Place the sensor head (15) flat onto the object to be measured. The sensor may not be moved or lifted dur-
ing measuring.
• Briefly press the measuring button (4 or 16) once. Measuring takes place automatically, which is signalled
with a beep. The display indicates „Mesu“.
To switch to HSL mode, press the button „RGB/HSL“. The display switches to this mode. Another press
returns the device to RGB mode.
• After measuring, turn the device off via the „POWER“ button.
REL comparative measurement
Via the REL function, you can perform a simple comparative measuring operation from a reference surface to
a sample surface. Deviating values can be easily detected as „+/0/-“ value.
Please proceed as follows to activate this function:
• Perform normal colour measurement on the reference object as described for RGB or HSL measuring.
• Once the measuring value is displayed, press the button „REL“ (8). The displayed values are saved as refer-
ence. The display blinks and shows „0 0 0“.
Place the sensor head on the sample surface and press the measuring button once. The comparison mea-
surement is in progress and shows the difference value after a brief measuring period.
• Deactivate the REL function by once again pressing the „REL“ button (8).