17
3. The meter turns on after a beep. An image of a test
strip with a flashing blood drop will appear letting you
know the meter is ready to test.
4. Shake the control solution bottle
thoroughly. Squeeze the control solution
bottle gently and discard the first drop.
Squeeze out a second small drop on a
clean nonabsorbent surface.
Note:
Do not apply control solution to the test strip directly
from the bottle.
5. Immediately touch the tip of the test strip to the
drop of control solution. The control solution is
pulled into the test strip through the strip tip.
Note:
If the control solution sample does not fill the check window, do not add a second
drop. Discard the test strip and start over with a new test strip.
6. Hold it in the drop until the meter beeps, and then you see the meter count
down on the screen and followed with your control test result after a beep.