Users Manual

16
Testing with Control Solution
Why Perform Control Tests
Performing a control test lets you know that your meter and test strips are working
properly to give reliable test results. You should perform a control test when:
At least once a week
You open a new box of test strip
You want to check the meter and test strips
Your test strips were stored in extreme temperature or humidity
After cleaning your meter
You dropped the meter
Your test result does not match with how you feel
About the Control Solutions
Only use VivaChek™ Eco Control Solutions (Low, Normal or High) to
practice on the system.
Your meter automatically recognizes the control solution.
The control solution results are not included in the average value
calculation.
All expiration dates are printed in Year-Month format. 2017-01 indicates
January, 2017.
Do not use control solution that is past the expiry date or discard date (the
control solution will expire 6 months after the bottle is opened for the first
time).
Shake the bottle well before use.
Close the bottle tightly after use.
Performing a Control Test
1. Remove a test strip from the test strip vial or foil pouch. Tightly close the vial
cap immediately after you have removed the test strip.
Note: Check the expiration and discard dates of the test strips. Do not use the
expiredtest strip.
2. Insert a test strip into the meter in the
direction of the arrows.