Reference:
1. ADA Clinical Practice Recommendations, 2014.
Warning:
• If your blood glucose reading is under 50 mg/dL (2.8 mmol/L) or you see LO
(less than 10 mg/dL (0.6 mmol/L)) on the meter display, contact your health
care professional as soon as possible.
• If your test result is above 250 mg/dL (13.9 mmol/L) or you see HI (greater
than 600 mg/dL (33.3 mmol/L)) on the meter display, contact your health
care professional as soon as possible.
• Do not change your medication therapy based on VivaChek™ Eco BLE test
result before consult your health care provider.
Questionable or Inconsistent Results:
If your blood glucose result does not match how you feel, please:
• Check the expiration date and the discard date of the test strip. Make sure
that the test strip vial has not been opened for more than 6 months.
• Confirm the temperature in which you are testing is between 5 and 45°C
(41-113°F).
• Make sure that the test strip vial has been tightly capped.
• Make sure the test strip has been stored in cool, dry place.
• Make sure the test strip was used immediately after removing from the test
strip vial or foil pouch.
• Make sure that you followed the test procedure correctly.
• Perform a control solution test (See Performing a Control Test on page13 for
instructions).
After checking all of the conditions listed above, repeat the test with a new test strip.
If you are still unsure of the problem, please contact your local dealer.
Removing the Used Lancet
Unscrew the lancing device cover. Place the
safety tab of the lancet on a hard surface and
carefully insert the lancet needle into the safety
tab.