Specifications
Section 8 RADIATION SAFETY
EQUIPMENT
PERSONAL MONITORING
ALARMS
Product Code Description
NCA0015 SM3 Personal Radiation Alarm
NCA0008 NDS Model ND15 Personal Radiation Alarm
NCA0009 NDS Model RA500 Personal Radiation Alarm
DOSIMETERS
Product Code Description
NCD0003 Arrow-Tech Mod 138 Quartz Fibre Pocket Dosimeter. Range 0 –200mR
NCD0011 Siemens Model EPD Mk2 Digital Dosimeter
NCD0012 MPG Mod DMC2000S Electronic Pocket Dosimeter
Introducing the new MPG Mod DMC 2000 S
Personal Electronic Dosimeter
Page 21 of 48
The DMC 2000S features flat energy response to X-
rays and Gamma field from 50keV to 6Mev and
linear response to dose rate fields from natural
background up to more than 10Sv/hr. The pass-by
data exchange feature gives unequal operational
flexibility
Dose and Dose rate
LCD Display
User friendly - Lightweight - Waterproof
Stand alone device or integrated into a Dosimetry
System
Audible and Visual Alarms
Large internal Histogram memory
Self-testing (Battery, Detector & Parameters
Hands free, pass-by exchange
Optional Teledosimetry or use as an Area Monitor
Physical Characteristics
Compliant to IEC 1283, ANSI 4220A
PTB approved
Display units: mSv or mrem
Measurement range:
o Dose: 1µSv to 10Sv (0.1mrem to 10rem)
o Rate: 0.1µSv/Hr to 10Sv/hr (0.01mrem/hr to1000
rem/hr
o Display: 0.01mSv/hr to 10 Sv/hr or 0.001mSv/hr to
10Sv/hr (extended option)
Linearity:
o <± 10% up to 1Sv/hr (100rem/hr)
o <± 25% up to 10Sv/hr (1000rem/hr)
X and Gamma energy range: 50keV to 6Mev
Accuracy: <± 5% (Cs137, 0.2mSv/hr)
Electrical Characteristics
Standard calculator battery LiMnO
2
CR2450 one year
battery life
Mechanical Characteristics
Dimensions: 84 x 48 x 17.5mm
Weight with battery: < 70g
Worn by a replaceable clip
Environmental Characteristics
Temperature range: -10ºC to 50ºC
Humidity: <90% at 42ºC
Storage: -30ºC to 71ºC
Shock, vibration and drop resistant, waterproof
EMC: compiles and exceeds standards by a large
margin
Customisation
Setup can be accessed by authorised user with
DOSIMASS software.
Back to Contents