4. Adjust gain
Connect the shunt resistor simulating strain desired for full scale. Adjust the COARSE
GAIN and FINE gain potentiometers for the desired full scale output.
All wire connections and potentiometers of this equipment are located on the surface of
the enclosure. No discrete gain setting is available, or any gain indications.
4.1.3.4 Comparison of Set-Up Procedures
Due to its lack of discrete gain setting mechanism and gain indication, and long set-up
procedure, Omega is less suitable for the target laboratories. Honeywell and Tacuna are
relatively easy to hook up and calibrate. For both systems, it is recommended that the
connections to be extended to avoid wear of the connection pieces and reduce issues caused by
bad connections in lab.
Comparing Honeywell UV-10 and Tacuna, Tacuna is more preferable. The students
won’t be required to open the enclosure of Tacuna if the gain setting is preset prior to the
laboratory, while the students need to open the enclosure of Honeywell UV-10 to adjust the
ZERO potentiometers, which might induce more complexity to managing the laboratories.
Tacuna also includes internal Wheatstone bridge completion circuit, in additional to the external
circuit set-up option tested in this project.